Valence region measured at 0.1 mbar O2 at various temperatures. The measurement was performed with an R3000 HP hemispherical, near-ambient pressure analyzer.
Ambient Pressure UV photoelectron spectroscopy (UPS) uses photons in the ultraviolet energy range, typically He I (21.2 eV) or He II (40.8 eV), to extract electrons from the valence bands of a sample at pressures in the mbar regime.
Ambient Pressure UV photoelectron spectroscopy (UPS) uses photons in the ultraviolet energy range, typically He I (21.2 eV) or He II (40.8 eV), to extract electrons from the valence bands of a sample at pressures in the mbar regime. In contrast to ‘regular’ UHV ultraviolet photoelectron spectroscopy, electron analyzers have to be adapted with differential pumping stages to be able to cope with the higher pressures in the setup. Beyond these more demanding requirements on the measurement hardware, ambient pressure UPS reveals the same kind of information as UHV UPS: high-resolution, high signal to noise spectra of the electronic valence states of the examined samples.
We use cookies to personalise content and ads, to provide social media features and to analyse our traffic. We also share information about your use of our site with our social media, advertising and analytics partners who may combine it with other information that you’ve provided to them or that they’ve collected from your use of their services